Tin offers high
conductivity at an affordable price.
9mm flat electrodes and
a 1.5mm (.059) DIN safety plug.
The special designed
safety plugs offer more space between safety plugs even at narrowly placed
sockets and a better fit also with wider (worn out) sockets at some head boxes
IMA electrodes with
Teflon wiring have 144 silver strands per lead.
IMA electrodes with
silicone coated wiring have 7 silver ribbons combined with Kevlar for strength,
dipped in silicone coating.
Each type : 48” = 1 ohm
resistance out of 4,000 ohm from scalp.
Care:
Do not soak in water
(creates corrosive brine).
Wipe with alcohol after
each use.
For heavy discoloration
or build up on silver electrodes use toothpaste (not gel), comet or ajax, wipe
gently and rinse with warm water.
For gold plated
electrodes use alcohol or toothpaste (not gel), be very gentle, rinse only - do
not soak. Scratched plating is the most common cause of failure for gold
electrodes.
Take care not to crimp
the wires or pull at wires from the ends, this is the most common reason of
failure for all electrode wiring.